My practical experience with endotracheal tubes...
Always maintain all sizes of endotracheal tubes for infants, small children and adult. Few adult cuffed endotracheal tubes are a must. Nasal soft tubes size 8, 9, 10 can be lifesavers. They should be sufficiently long. They can also be used for passing a Ryle’s tube in an unconscious patient. After threading the Ryle’s tube in a well-lubricated nasal tube is passed in Oesophagus and Ryle’s tube is threaded down into the oesophagus, the nasal tube is withdrawn. If it is difficult to pass the tube nasally, pass the tube orally push the threaded Ryle’s tube and withdraw the soft tube. The oral Ryle’s tube can be railroaded into the nose with the help of catheter fixed snugly in the end of Ryle’s tube. Push the Ryle’s tube, which is stiff. Do not pull the catheter.
In case of difficult intubation a soft portex tube is passed nasally which mostly enters the larynx. Sometimes it remains anterior to the epiglottis, which can be placed posterior to the epiglottis with Magill’s forceps and threaded to enter the larynx. Sometimes the tube gets hitched against the anterior commissure and cannot advance in the larynx and trachea. It can be manoeuvred by slightly flexing the neck, which corrects the angle between the advancing tube and the trachea.
All endotracheal tubes must be washed by you with soap and water and a jet of water forced down the tube to remove all the left over secretions and finally cleaned by brush and water and dried with a towel. This will save your embarrassment when a freshly boiled tube for use shows secretions of the previous case or a foreign body like a cotton swab or even a cockroach, even a nasal turbinate has been recovered.
Nasal endotracheal tube advanced before oral extubation in case of full stomach makes life easier. Sometimes solid food and big blood clots cannot be aspirated from the oropharynx when extubated and the patient struggle for breath. Nasal tube maintains the free airway and allows sufficient time for the patient to vomit out stomach contents if he is conscious and clear the oropharynx.
All tubes should be fitted with female Noseworthy’s connections. Cobb’s connection is only useful in cases with wet lungs when intermittent suction is mandatory. The connection must be tight fitting and of correct size. Do not try to fit bigger connection as it may tear the tube (rubber).
Cuffed ET should not be used all the time. It spares your cuffed tubes. With use of air it is hardly necessary if the bellows or respirator are used. As far as possible do not inflate the cuff. Pass the proper size tube. If the air leaks by the side of tube is more, replace the tube with the next size. Use of cuff can cause oedema and damage to the mucous membrane of the trachea in long duration cases. After placement of ET inflate the cuff yourself. Over zealous Assistants always tend to over inflate the cuff. The over inflated cuff becomes a very strong stimulus and in light plane of anaesthesia it can lead to intense bronchospasm endangering life if not detected. Cases are known where inflation becomes impossible due to bronchospasm and cyanosis start appearing in spite of oxygen rich mixture used to inflate the lungs thus produced by an inflated balloon. The bronchospasm produced by over inflated balloon does not respond to usual bronchodilator treatment like injection of drugs. It only responds to:
-Deflation of cuff
-Removing the ET
-Deepening the level of anaesthesia
The simplest in the situation is deflation of cuff. The remaining methods appear to be dangerous and are not productive. For conduction of cases always have 3 sizes ready. Also do not forget to keep a nasal tube of appropriate size always ready. It is specially important in cases who are overweight, short neck or where the mouth cannot open adequately. Nasal intubation can prove a life saving procedure in all difficult intubations. In a planned case always ascertain the freedom of breathing through each nostril separately. Lubricate the nasal tube thoroughly, do not use force at any stage. You can have two obstructions while passing a tube through the nose, one at the nasopharynx where it meets the Passavant’s ridge and secondly while the tube meets the larynx at its outlet. The first difficulty should be solved by putting a finger in the nasopharynx as you advance the tube to hook it and bring it out into the pharynx, and second difficulty can always be solved by flexing the neck so that the angle of the advancing tube corresponds with the trachea. The difficulty is due to the nasal tube hitching against the anterior commissure. Do not force the nasal tube. It can cause bleeding.
Rubber tubes last about a year or more and hence one should not hesitate to boil the tube for 2 minutes before each case. Discard the tube when the tube shows signs of perish. The boiled tubes before use must be cooled and in emergency cooled under tap water. A case of an adult female is recorded who had an intubation with hot tube. This patient for Thyroidectomy needed Tracheostomy in the postoperative period. This precaution is a must in case of children and infants.
Inspect the tube before giving it for boiling for any foreign body (Cockroach) or debris. Sometimes cotton swabs used for cleaning have been detected. So look down the tube after straightening to find the presence of any obstruction. Unimaginable objects have been accidentally found inside the tube. In small tubes and Oxford tubes run the water through the tube and pass a stillette to ensure perfect patency. One must possess all types of tubes like armoured, nasal, Oxford, infant and special for specific use. If one can obtain laryngeal mask airway all 3 sizes.
Endotracheal tube connectors
One should have sets of different connections like right angle, Cobb’s, and curved as in some situations one requires various angles and various other uses for the connector. A Cobb’s connection can be used when one has to resort to suction intermittently and for giving IPPR with thumb technique and for use as Ayre’s T piece if it is not available. For routine use it is better to have a plastic female nose worthy connection fitted to each tube and a male Noseworthy connection to the catheter mount or the various unidirectional valves can directly connected to Noseworthy connection with metal connectors check the catheter mount.
Comments?
Wednesday, March 25, 2009
Endotracheal Tube
Tuesday, March 24, 2009
Your equipment is handled by you alone and no one can help you
Make sure that your equipment is full and complete and you are capable of conducting any case even in a village. You must be fully equipped with all your drugs for induction, maintenance, reversal, and also for resuscitation. Do not depend upon the nursing home to provide you with anything.
One of my colleagues never carried distilled water vials with him. One day he was faced with no saline, no water to dissolve Thiopentone. He probably used boiling water from the sterilizer and no wonder all patients developed hyperpyrexia.
When you paralyze a patient with long acting muscle relaxant you must possess the necessary prostigmine with you. Similarly if you intend to give spinal a vasopressor must be in your possession. While anaesthetizing an asthmatic patient check the immediate availability of aminophylline, cortisone, adrenalin etc. and the spray he may be using (salbutamide).
Laryngoscope
Laryngoscope is the most important piece of equipment. It must be in perfect condition. All spares must be readily available all the time. A spare bulb and pair of batteries (unused) must be at hand. Those who use rechargeable batteries, a note of caution, when the batteries are nearly discharged the bulb can fade away within seconds. A pair of batteries must always be in stock. The repairs and checking of all parts must be mastered for use in emergency. If the bulb is not working, check the following:
Change the batteries.
Check all contacts points and clean all the metal contacts with rough hard object or sandpaper if available.
Check the bulb by unscrewing from its place and checking it on the spring of the metal contact on laryngoscope handle and touching the body of the bulb to the body of the handle. If the bulb lights it means that laryngoscope handle, batteries, and bulb are O.K.
Clean the seating of the bulb in the blade and re-check the contact point on the blade and polish the same with sandpaper or even a metal file. The seating of the bulb can be cleaned by a needle and a drop of ether. Allow it to dry and replace the bulb.
Check the bulb. If it shows blackish discoloration change the bulb.
The insulation between the bulb body and contact points is only a small piece of paper, which gets wet while cleaning with water. Disengage the bulb and let it dry. Try to use it again. If it does not, pull the metal contact by a fraction of a millimetre. After constant use the insulation (paper) becomes useless and the contact and the bulb body touch directly and the bulb is short-circuited.
Checking the handle is easy. Unscrew the cap of the handle. Check the tension of the spring. Dry the handle from inside. Polish the contact of the spring with the metal body of the battery. Thus all metal contact points must be checked and cleaned of deposition of copper oxides. After use detach the blade, screw the bulb (if it is loose) and wash it with soap and water. The handle does not need cleaning unless smeared with blood during use in a difficult intubation. Minor points in repair of the laryngoscope must be known. Nobody teaches nor is it given in books. Check the light after cleaning.
If possible keep a spare Laryngoscope, cells and a bulb.
Extra long blade and special laryngoscopes can be useful.
Lastly do not loose your head. Insert the un-lighted blade in the patient’s mouth, give the correct position for intubation and pass a tube blindly. If the intubation for the same reason is difficult you may fail, otherwise in 90% cases you will succeed and carry the day.
Endotracheal tubes
The length of the endotracheal tube (ET) must not be longer than 2” of the size required. Longer ET may cause endobronchial intubation. The ET must have a nose worthy connection fitted permanently. So that it can be easily connected to the unidirectional valve for IPPR. Longer ET comes in the way for operations on the head and face and intra-oral surgery.
All tubes must be boiled for 2–3 minutes before use and re-checked as some debris only becomes apparent after boiling. The hard plastic tubes should be kept in warm water to make them soft and non traumatic (for nasal intubation).
Always check the temperature of the ET after it is taken out from the boiling water. Allow it to cool. Apply lubricant jelly after the tubes have cooled. In extreme emergency cool the tubes under tap water before intubating.
Keep all sizes of ET handy all times including OO size for new born resuscitation. Always have 3 sizes of ET ready. Apart from the proper size that you have selected have a smaller size and bigger size ready. Do not fumble at the last moment.
Special nasal tubes of sizes 6 mm to 10 mm must always be available for emergency use in failed oral intubation.
All endotracheal tubes must be secured with a good strip of sticking plaster. In cases of patient with beard or patient in prone position, it must be secured with additional bandage going round the neck. Chances of slipping of tube (which are quite high) are thus prevented. In case of infants it can be a disaster. In cases of children never pass a cuffed endotracheal tube. Avoid using cuffed tubes in cold cases. If at all used, inflate the cuff balloon yourself.
Other equipment
All tubes should be fitted with Noseworthy connections of proper size so as to avoid searching for a connection and fitting of wrong connection (small) in a bigger endotracheal tube (by the assistant or ward boy) the Nose worthy connection takes up the unidirectional valve directly and does not need any other connection or catheter mount. In some cases of facio maxillary surgery (cleft lip and palate) use the Magill’s connection going to the endotracheal tube (especially the Oxford tube) with extra length of arm going out. This makes the job of the plastic surgeon of fitting the Boyle Davis gag easy and leaving some length of connection for fitting Ayre’s T piece.
Bellows Unit: Always open the bellows from the lock instead of fumbling. Never keep cotton swabs and gauze pieces on the table as they are likely to get sucked in and block the inlet of the bellows and after a time may not allow any air in the bellows. These foreign bodies are not easily spotted unless you expect them. If the rubber shows signs of perishing get it changed immediately.
It is always preferable to use the bellow unit with unidirectional valve situated under the bellows than separate disc valve set. Like in Oxford inflating bellows. It makes the equipment lighter.
Disconnect the delivery tube whenever one leaves the bottle or the patient. Always keep the palm loosely over the lid which keeps you informed about the respiratory rate and volume and it steadies the vaporizer. So many accidents have been recorded due to aspiration of pure liquid ether in the respiratory passage which results in pulmonary oedema and death. At one time this technique was supposed to be the safest. However if taken lightly it can cause deaths.